Chapter 5—Alaris™ SpO
2
Module Models 8210 and 8220
278 BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual
Models 8210 and 8220
Feature Default Setting Options
Limit Mode Adult Adult, Neonatal
Pulse Beep Volume 1 1, 2, 3, Off
Pulse Rate Alarm Limit, High Adult Mode: 120 bpm
Neonatal Mode: 200 bpm
31 - 240 bpm
Pulse Rate Alarm Limit, Low Adult Mode: 50 bpm
Neonatal Mode: 100 bpm
30 - 239 bpm
SpO
2
Alarm Limit, High Adult: Off
Neonatal: 95%
21 - 100%, Off
SpO
2
Alarm Limit, Low Adult: 90%
Neonatal: 80%
20 - 99%
Model 8210
Feature Default Setting Options
SatSeconds Off 10, 25, 50, 100 seconds; Off
Model 8220
Feature Default Setting Options
Saturation Averaging Time
(display update period)
8 seconds 2, 4, 8, 10, 12, 14, 16 seconds
Sensitivity Mode Normal Normal, Maximum