Chapter 5Alaris™ SpO
2
Module Models 8210 and 8220
280 BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual
NOTE:
Wavelength range information can be especially useful to clinicians, f
or example, those performing
photodynamic therapy.
Model 8210
Accuracy Tolerance:
Display Update Period:
2.25 seconds
Measurement Range:
Perfusion: 0.03 - 20%
Pulse Rate: 20 - 250 bpm
SpO
2
: 1 - 100%
Pulse Amplitude Display:
Visual indicators for pulse signals represent proportional pulse amplitude strength.
Sensor:
Emitted light wavelength range is within 500 - 1000 nm. Output power does not
exceed 15 mw.
Model 8220
Accuracy and Motion
Tolerance:
Display Update
Period:
Approximately
1
second.
Measurement Range:
Perfusion:
0.02
-
20%
Pulse Rate:
25
-
240
bpm
S
p
O
2
:
1
-
100%
Pulse Amplitude
Display:
Proportional to height of
I.Q.
signal.
Sensor:
Emitted light wavelength range is within
500
-
1000
nm. Output power does not exceed
1
mw.
Low Perfusion
Adult
Neonate
Pulse Rate 20 - 250 bpm
±3 digits
20 - 250 bpm
±3 digits
20 - 250 bpm
±3 digits
Functional
Saturation
70 - 100%
±2 digits
70 - 100%
±2 digits
70 - 100%
±3 digits
Low
Perfusion
Motion

No Motion
Resolution
Pulse Rate 25 - 240 bpm
Adults,
Pediatrics,
Neonates: ±3
digits
25 - 240 bpm
Adults,
Pediatrics,
Neonates: ±5
digits
25 - 240 bpm
Adults,
Pediatrics,
Neonates: ±3
digits
1 bpm
Saturation 70 - 100%
Adults,
Pediatrics: ±2
digits;
Neonates: ±3
digits
70 - 100%
Adults,
Pediatrics,
Neonates: ±3
digits
70 - 100%
Adults,
Pediatrics: ±2
digits;
Neonates: ±3
digits
1% SpO