Programming with Interoperability and Guardrails™ Suite MX
BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual 141
Changing Primary Infusion Parameter
1. Press the CHANNEL SELECT key.
2. Press the
SET UP soft key.
3. Press the
PRIMARY soft key.
4. To change primary infusion parameter, press the applicable soft key (
RATE or VTBI) and use numeric
data entry keys.
5. Verify correct primary infusion parameters and press the
SECONDARY soft key.
Secondary setup screen is displayed.
6. Press the
NEXT soft key.
7. To resume secondary infusion, press the
START soft key.
8. Verify that drops are flowing from the secondary container drip chamber.
Stopping Secondary and Returning to Primary
1. Press the CHANNEL SELECT key.
2. Press the
SETUP soft key.
3. Press the
PRIMARY soft key.
4. Close the clamp on the secondary administration set.
or
Disconnect the secondary administration set from the upper inj
ection port.
5. Press the
START soft key.
6. To stop the secondary infusion and begin infusing the primary, press the
Yes soft key.
The secondary infusion stops and primary infusion begins. The main screen appe
ars.
NOTE:
The SEC to PRI alert does not sound when th
e infusion is manually ended and returned to the primary.