6. Close the door and latch, as follows:
a. Close the door and hold it in a closed position by grasping both the door and the instrument case
with o
ne hand.
b. Gently lower the latch.
The safety clamp device is automatically
dis
engaged.
WARNING
Do not touch the administration set while closing the door. Failure to follow
this instruction can result in infusion rate inaccuracy.
Ensure tubing placement is over pressure sensors.
Ensure the upper fitment is not elevated above the receptacle on the pump.
This could be caused by maintaining traction on the set (or stretching the
set) while closing and latching the door. If the set is loaded in this manner,
infusion inaccuracies may occur. See example and pictures.
To prevent a potential free-flow condition, ensure that no extraneous object
(for example, bedding, tubing, glove) is enclosed or caught in the pump
module door.