Chapter 3BD Alaris™ Pump Module Model 8100, Alaris™ Pump Module Model 8100, and Alaris™
Syringe Module Model 8110
122 BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual
The following procedures are to be used only when the drug to be infused is listed in the drug library. To
access the drug library, a hospital-defined best-practice data set must be transferred using the Guardrails
editor software and the profiles feature must be enabled.
NOTE:
There is a potential for a discrepancy in
precision between the numeric values displayed on the system
(PCU, pump module and syringe module) and in the electronic medical record (EMR)/hospital information
system (HIS) due to different rounding rules. The device accepts a limited set of precision values for
manual input and display, however, all program values are calculated and reported to 4 decimal places of
precision.
NOTE:
There is a potential that the concentration displayed on the PCU will use different units of measure than the
concentration displayed in the EMR/HIS. The system PCU displays the concentration (drug amount per
1 mL) using the dosing units assigned to tha
t Guardrails drug entry. (Note: The system allows Continuous
Infusions to be dosed in different units than the units used in the Drug Amount.) However, the system
sends the concentration to the EMR/HIS using the units of the Drug Amount.
It is important to note tha
t the concentration displayed on the PCU should be equivalent to the
concentration displayed in the EMR/HIS even if different units of measure are used.
The following example is for a continuous drug that has a Drug Amount in grams, but is dosed in mg:
NOTE:
It is possible to program an infusion with a rate that is display
ed with
two decimal places (one-hundredth of
a mL per hour) on the PCU for the pump module. However, due to space limitations on the pump module
rate display, the rate is displayed to the nearest one-tenth of a mL per hour on the pump module. T
his
value is only used for display purposes and the pump module is actually infusing at the m
ore precise
rate noted on the PCU.
NOTE:
If the infusion order does not specify whether the infusion is to b
e pre-populated as a primary or secondary
infusion, the system will determine whether to pre-populate the infusion as primary or secondary based on
the state of the intended module and on the Guardrails drug library configuration. For example: The
infusion order will be pre-populated as a primary infusion on an idle pump module if the configuration of the
drug/fluid entry in the active Guardrails profile supports it. The infusion order will be pre-populated as a
secondary infusion on a pump module that is programmed for or infusing a primary fluid if the
configurations of both the primary and secondary entries in the active Guardrails profile support it.
NOTE:
All fields that are editable with manual pr
ogramming are editable when pre-populated with interoperability.
Displayed on PCU Infusion information sent
to EMR/HIS
Drug
Amount: 1 gram
Diluent Volume: 100 mL
[Conc]: 10 mg/mL
Drug Amount: 1 gram
Diluent Volume: 100 mL
[Conc]: 0.01 gram/mL

Table of Contents