Chapter 3—BD Alaris™ Pump Module Model 8100, Alaris™ Pump Module Model 8100, and Alaris™
Syringe Module Model 8110
190 BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual
NOTE:
The minimum Volume To Be Infused (VTBI) for both pum
p and syringe modules (applicable to all syringe
sizes) is 0.1mLs. This applies to Continuous Infusions, bolus dose, Intermittent Infusions, and IV Fluids.
The set length and temperature of the administration set can affect occlusion time to alarm. Increases in
temperature and increases in length may increase time to alarm.
Time to Alarm,
Maximum:
Volu
me to be Infused
Programming
Increments:
Weight: 2.6 lbs
Pressure Limit (mmHg) Rate (mL/h) Time to Alarm
50 0.1
≤
50 minutes
50 1
≤
5 minutes
50 25
≤1
5 seconds
525 0.1
≤
7.5 hours
525 1
≤4
5 minutes
525 25
≤
2 minutes
Range (mL) Increments (mL)
0.1 - 9.99 0.01
10 - 999.9 0.1
1000 - 9999 1