Chapter 4Alaris™ PCA Module Model 8120
250 BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual
NOTE:
This configuration setting is a shared setting between the PCA module and the syringe module.
These values are configured in the SpO
2
module settings within the Guardrails editor software and can
be changed by the clinician by accessing Channel Options on the PCA module.
These values are configured in the EtCO
2
module settings within the Guardrails editor software and can
be changed by the clinician by accessing Channel Options on the PCA module.
PCA Pause Protocol:
PCA Pause Protocol
Monitoring Module Attach
Enforcement
PCA Pause Protocol Text
SpO2 Settings:
% SpO
2
Low Limit
Initial Value
EtCO
2
Settings:
Respiratory Rate Lower Limit
(bpm)
Initial Value
Disabled
None
PCA infusion has paused due to
a decline in respiratory status.
Check patient.
None
None
None
None
Enabled - Disabled
Enabled - Disabled
Editable per hospital protocol
20 - 99
20 - 99
0 - 149
0 - 149
Priming
Enabled Enabled- Disabled
Security Access Level
Level 1 Level 1, 2, 3
Feature Default Setting Options