Chapter 5Alaris™ SpO
2
Module Models 8210 and 8220
282 BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual
The loss of a pulse signal can occur in any of the following situations:
The sensor is too tight.
Exposure to excessive illumination; such
as, a surgical lamp (especially one with a xenon light source),
bilirubin lamp, fluorescent light, infrared heating lamp, or direct sunlight.
The sensor placed on an extremity with a blood pressure cuff, arterial catheter
, intravascular line, or
other causes of insufficient perfusion.
The patient has hypotension, severe vas
oconstriction, severe anemia or hypothermia, is in cardiac
arrest or is in shock.
There is an arterial occlusion proximal to sensor.
Placement is too close to electrosurgery e
quipment.
NOTE:
Exposure to excessive illumination can be corrected by covering the sensor with a dark or opaque
material.