Appendix A—Troubleshooting and Maintenance
334 BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual
DELAY COMPLETE Syringe
Pump
Medium Delay time complete. To resume the infusion,
press
RESTART
on the
module or press
CHANNEL SELECT
and
then the
START
soft key.
DELAYED STATUS Syringe
Pump
Informational The infusion has been delayed
for a defined
period of time.
No user response is
expected.
DISCONNECT
OCCLUDED
DISPOSABLE
EtCO
2
Medium The disposable is occluded. Check the disposable.
Obtain a new disposable
and attach it to patient.
DISPLAY FAILURE -
REPLACE
MODULE
SpO
2
EtCO
2
High There is a malfunction with the
display on the SpO
2
or EtCO
2
modules. The device continues
to monitor.
Replace the module.
DISPOSABLE
DISCONNEC
TED
EtCO
2
Informational No disposable is connected to
the module while attempting to
search for a patient.
Ensure disposable is
attached to module.
DISPOSABLE
DISCONNEC
TED
EtCO
2
Medium No disposable is connected to
the module during monitoring or
after approximately 30 seconds
of searching for a patient.
Ensure disposable is
attached to module.
HIGH ETCO2 EtCO
2
High The EtCO
2
value is above the
specified limit.
Assess the patient's
condition. Confirm that
the correct alarm limit
values are selected.
HIGH FICO2 EtCO
2
High FiCO
2
value is above the
specified limit.
Assess the patient's
condition. Confirm that
the correct alarm limit
values are selected.
HIGH RR EtCO
2
High The respiratory rate is above the
specified limit.
Assess patient's the
condition. Confirm that
the correct alarm limit
values are selected.
HIGH SATURATION SpO
2
High The high saturation alarm limits
have been exceeded.
Assess the patient's
condition. Confirm that
the correct alarm limit
values are selected.
HIGH_PULSE SpO
2
High The high pulse rate alarm limit
has been exceeded.
Assess the patient's
condition. Confirm that
the correct alarm limit
values are selected.
INCOMPATIBLE
MODULE
PCU In
formational The PCU is unable to recognize
the attached MODULE because
it is an unsupported module type
Remove the unsupported
module.
INFUSION
COM
PLETE
Syringe
Pump
Medium The current infusion is
completed with no KVO.
Set up a new infusion or
press the
CHANNEL
OFF
key.
INFUSION
COM
PLETE - KVO
Pump
Syringe
Medium
VTBI has been infused; the
module is i
nfusing at the KVO
rate
Set up a new infusion, if
required, and reinititate
the infusion
Alarm Message Device Priority Meaning Response