Regulations and Standards
BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual 357
Table 2 (Continued)
Electromagnetic Immunity
Immunity Test
IEC 60601-1-
2:2014
Test Level
Compliance
Level
Electromagnetic Environment -
Guidance

IEC 61000-4-6
Conducted RF
IEC 61000-4-3
Radiated RF
IEC 61000-4-3
Proximity fields
from RF wireless
communication
equipment
3Vrms 0.15 MHz
- 80MHz
6Vrms in ISM
bands between 0.15
MHz and 80 MHz
80% AM at 1 kHz
3V/m
80 MHz -
2.7 GHz
80% AM at 1 kHz
See Table 9 in
IEC 60601-1-2:2014
3 Vrms, 0.15
MHz - 80MHz
6Vrms in ISM
bands between 0.15
MHz and 80 MHz
80% AM at 1 kHz
3V/m
80MHz - 2.7 GHz
80% AM at 1 kHz
See Table 9 in
IEC 60601-1-2:2014
Portable and mobile RF communications equipment
sh
ould be used
no closer to Alaris System
(including cables) tha
n recommended separation
distance calculated from equation applicable to
frequency of transmitter.
Recommended Separation Distance:
12
__ d = [-----] √P
V
2
12
__ d = [-----] √P 80 MHz - 800 MHz
E
1
12
__ d = [-----] √P 80 MHz - 2.5 GHz
E
1
d = recommended separation distance in meters
(m).
P = maximum output power rating of transmitter in
watts (W) according to transmitter manufacturer.
Field strengths from fixed RF transmitters, as
determined
by an electromagnetic site survey,
should be less than compliance level in each
frequency range.

Interference might occur in vicinity of equipment
marked with the following symbol: