Model 2450
Interactive SourceMeter® Instrument Reference Manual Section 2:
General operation
2450-901-01 Rev. B/September 2013 2-73
Setting the instrument to use the front or rear terminals
If the output is on when you change the settings for the terminals that are used, the output turns off.
Using the front panel:
1. Press the FUNCTION key.
2. Select the source and measurement combination
3. Press the TERMINALS FRONT/REAR switch.
When F is lit, the instrument is using the front-panel terminals. When R is lit, the instrument is using
rear-panel terminals.
Using SCPI commands:
To change to the front-panel terminals for current measurements, send the command:
ROUTe:TERMinals FRONt
To change to the rear-panel terminals for current measurements, send the command:
ROUTe:TERMinals REAR
Using TSP commands:
To change to the front-panel terminals, send the command:
smu.measure.terminals = smu.TERMINALS_FRONT
To change to the rear-panel terminals, send the command:
smu.measure.terminals = smu.TERMINALS_REAR

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
BrandKeithley
Model2450
CategoryTest Equipment
LanguageEnglish

Related product manuals