Model 2450
Interactive SourceMeter® Instrument Reference Manual Section 6:
SCPI command reference
2450-901-01 Rev. B/September 2013 6-59
[:SENSe[1]]:<function>:UNIT
This command describes the units of measurement that are displayed on the front panel of the instrument and
stored in the reading buffer.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
Current: AMP
Voltage: VOLT
Resistance: OHM
Power: WATT
Usage
[:SENSe[1]]:CURRent:UNIT <currentMeasure>
[:SENSe[1]]:CURRent:UNIT?
[:SENSe[1]]:VOLTage:UNIT <voltageMeasure>
[:SENSe[1]]:VOLTage:UNIT?
<currentMeasure>
OHM, WATT, or AMP
<voltageMeasure>
OHM, WATT, or VOLT
Details
The change in measurement units is displayed when the next measurement occurs.
Example
VOLT:UNIT WATT
Changes the front-panel display and buffer
readings for voltage measurements to be
displayed as power readings in watts.
Also see
None