EasyManuals Logo

Keithley 2450 User Manual

Keithley 2450
878 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #345 background imageLoading...
Page #345 background image
Section
6: SCPI command reference Model 2450 Interactive SourceMeter® Instrument
Reference Manual
6-28 2450-901-01 Rev. B/September 2013
FORMat
The commands for this subsystem select the data format that is used to transfer instrument readings
over the remote interface.
:FORMat:ASCii:PRECision
This command sets the precision (number of digits) for all numbers returned in the ASCII format.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
0
Usage
:FORMat:ASCii:PRECision <n>
:FORMat:ASCii:PRECision?
:FORMat:ASCii:PRECision? MINimum
:FORMat:ASCii:PRECision? MAXimum
:FORMat:ASCii:PRECision? DEFault
<n>
The precision:
Automatic: 0
Specific value: 1 to 16
MINimum (0)
MAXimum (16)
DEFault
(0)
Details
This attribute specifies the precision (number of digits) for queries.
Note that the precision is the number of significant digits. There is always one digit to the left of the decimal point;
be sure to include this digit when setting the precision.
Example
:FORM:ASC:PREC 10
Set a precision of 10 digits. An example of the
output is:
-6.999999881E-01
Also see
:FORMat[:DATA] (on page 6-30)

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
Model2450
CategorySourceMeter
Voltage Source Range±200 V
Current Source Range±1 A
DC Voltage200 V
DC Current1 A
Power Output20 W
Resolution6.5 digits
Display5-inch capacitive touchscreen
Operating Temperature0°C to 50°C
ConnectivityUSB, LAN, GPIB
Dimensions213 mm × 88 mm × 370 mm

Related product manuals