2: General operation Model 2450 Interactive SourceMeter® Instrument
Reference Manual
2-96 2450-901-01 Rev. B/September 2013
Offset-compensated ohms
The voltage offsets because of the presence of thermal EMFs (V
EMF
) can adversely affect resistance
measurement accuracy. To overcome these offset voltages, you can use offset-compensated ohms.
See Offset-compensated ohm calculations (on page 4-20
) for additional detail on calculating
offset-compensated ohms.
Setting offset-compensated ohms
Using the front panel:
This setting is only available from the front panel when the resistance measurement function is
selected.
1. Press the MENU key.
2. Under Measure, select Settings.
3. Next to Offset Comp, select On.
4. Select HOME to return to the operating display.
Using SCPI commands:
To enable offset-compensated ohms, send the command:
SENSe:RESistance:OCOMpensated ON
Using TSP commands:
To enable offset-compensated ohms, send the commands:
smu.measure.func = smu.FUNC_RESISTANCE
smu.measure.offsetcompensation = smu.ON
Source and measure using SCPI commands
The SCPI commands that set up the source functions are in the SOURce subsystem.
The source commands are specific to each source function (voltage or current). For example, to set
the range to 100 mA for the current function, you would send:
:SOURce:FUNCtion CURRent
:SOURce:CURRent:RANGe 1e-01
To set the range to 20 V for the voltage function, you would send:
:SOURce:FUNCtion VOLTage
:SOURce:VOLTage:RANGe 20
The SCPI commands that set up the measurement functions are in the SENSe subsystem.
The sense commands are also specific to each measure function (voltage, current, or resistance). For
example, to set the NPLC cycles to 0.5 for the current measurement function, you would send:
:SENSe:CURRent:NPLCycles .5
For the voltage measurement function, you would send:
:SENSe:VOLTage:NPLCycles .5
For the resistance measurement function, you would send:
:SENSe:RESistance:NPLCycles .5

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
Model2450
CategorySourceMeter
Voltage Source Range±200 V
Current Source Range±1 A
DC Voltage200 V
DC Current1 A
Power Output20 W
Resolution6.5 digits
Display5-inch capacitive touchscreen
Operating Temperature0°C to 50°C
ConnectivityUSB, LAN, GPIB
Dimensions213 mm × 88 mm × 370 mm