Interactive SourceMeter® Instrument Reference Manual Section 8:
2450-901-01 Rev. B/September 2013 8-167
Example
trigger.blender[1].orenable = true
trigger.blender[1].stimulus[1] = trigger.EVENT_DIGIO3
trigger.blender[1].stimulus[2] = trigger.EVENT_DIGIO5
Generate a trigger blender 1
event when a digital I/O
trigger happens on line 3 or
Also see
trigger.blender[N].reset() (on page 8-167)
trigger.blender[N].overrun
This attribute indicates whether or not an event was ignored because of the event detector state.
Type TSP-Link accessible Affected by Where saved Default value
Trigger blender N clear
Usage
overrun = trigger.blender[N].overrun
Trigger blender overrun state (true or false)
The blender number (1 or 2)
Details
Indicates if an event was ignored because the event detector was already in the detected state when the event
occurred. This is an indication of the state of the event detector that is built into the event blender itself.
This command does not indicate if an overrun occurred in any other part of the trigger model or in any other
trigger object that is monitoring the event. It also is not an indication of an action overrun.
Example
print(trigger.blender[1].overrun)
If an event was ignored, the output
is true.
If an event was not ignored, the
Also see
trigger.blender[N].reset() (on page 8-167)
trigger.blender[N].reset()
This function resets some of the trigger blender settings to their factory defaults.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
trigger.blender[N].reset()
The trigger event blender (1 or 2)