Interactive SourceMeter® Instrument Reference Manual Section 8:
TSP command reference
2450-901-01 Rev. B/September 2013 8-167
Example
trigger.blender[1].orenable = true
trigger.blender[1].stimulus[1] = trigger.EVENT_DIGIO3
trigger.blender[1].stimulus[2] = trigger.EVENT_DIGIO5
Generate a trigger blender 1
event when a digital I/O
trigger happens on line 3 or
5.
Also see
trigger.blender[N].reset() (on page 8-167)
trigger.blender[N].overrun
This attribute indicates whether or not an event was ignored because of the event detector state.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (R)
Yes
Instrument reset
Trigger blender N clear
Trigger blender N reset
Not applicable
Not applicable
Usage
overrun = trigger.blender[N].overrun
overrun
Trigger blender overrun state (true or false)
N
The blender number (1 or 2)
Details
Indicates if an event was ignored because the event detector was already in the detected state when the event
occurred. This is an indication of the state of the event detector that is built into the event blender itself.
This command does not indicate if an overrun occurred in any other part of the trigger model or in any other
trigger object that is monitoring the event. It also is not an indication of an action overrun.
Example
print(trigger.blender[1].overrun)
If an event was ignored, the output
is true.
If an event was not ignored, the
output is false.
Also see
trigger.blender[N].reset() (on page 8-167)
trigger.blender[N].reset()
This function resets some of the trigger blender settings to their factory defaults.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
trigger.blender[N].reset()
N
The trigger event blender (1 or 2)

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
Model2450
CategorySourceMeter
Voltage Source Range±200 V
Current Source Range±1 A
DC Voltage200 V
DC Current1 A
Power Output20 W
Resolution6.5 digits
Display5-inch capacitive touchscreen
Operating Temperature0°C to 50°C
ConnectivityUSB, LAN, GPIB
Dimensions213 mm × 88 mm × 370 mm