Model 2450
Interactive SourceMeter® Instrument Reference Manual Section 6:
SCPI command reference
2450-901-01 Rev. B/September 2013 6-93
:STATus:QUEStionable:ENABle
This command sets or reads the contents of the questionable event enable register of the status model.
Type Affected by Where saved Default value
Command and query
STATus:PRESet
Nonvolatile memory
0
Usage
:STATus:QUEStionable:ENABle <n>
:STATus:QUEStionable:ENABle?
<n>
The value of the register
Details
This command sets or reads the contents of the Enable register of the Questionable Event Register.
When one of these bits is set, when the corresponding bit in the Questionable Event Register or Questionable
Condition Register is set, the MSB and QSM bits in the Status Byte Register is set.
For detail on interpreting the value of a register, see Understanding bit settings (on page C-15
).
Example 1
:STAT:QUES:ENAB 8
:STAT:QUES:ENAB?
Enable bit 4, Limit 3 Fail, when the limit
test 3 failure value is exceeded. Check to
see that the value was set.
Also see
None
:STATus:QUEStionable[:EVENt]?
This command reads the Questionable Event Register.
Type Affected by Where saved Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:STATus:QUEStionable[:EVENt]?
Details
This query reads the contents of the questionable status event register. After sending this command and
addressing the Model 2450 to talk, a value is sent to the computer. This value indicates which bits in the
appropriate register are set.
The Questionable Register can be set to the numeric equivalent of the bit to set. To set more than one bit of the
register, set the Questionable Register to the sum of their decimal weights. For example, to set bits B12 and
B13, set the Questionable Register to 12,288 (which is the sum of 4,096 + 8,192).
Example
:STAT:QUES?
Query the Questionable Register.
Also see
Questionable Event Register (on page C-7)

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
Model2450
CategorySourceMeter
Voltage Source Range±200 V
Current Source Range±1 A
DC Voltage200 V
DC Current1 A
Power Output20 W
Resolution6.5 digits
Display5-inch capacitive touchscreen
Operating Temperature0°C to 50°C
ConnectivityUSB, LAN, GPIB
Dimensions213 mm × 88 mm × 370 mm