Model 2450
Interactive SourceMeter® Instrument Reference Manual Section 9:
Frequently asked questions (FAQs)
2450-901-01 Rev. B/September 2013 9-5
What to do if the GPIB controller is not recognized?
If the hardware is not recognized by the computer:
1. Uninstall the software drivers.
2. Reboot the computer.
3. Check for newer drivers on the vendor’s website. Check that the drivers are valid for the
operating system you have and any updates that might be necessary. This information is typically
found in the readme file that comes with the drivers.
4. Install software drivers.
5. Reboot the computer.
6. Plug in the hardware.
If it is still not recognized, you can try a different computer using a different operating system to rule
out operating system issues.
If this does not resolve the issue, contact the vendor of the GPIB controller for assistance.
I'm receiving GPIB timeout errors. What should I do?
If your GPIB controller is recognized by the operating system, but you get a timeout error when you
try to communicate with the instrument, check the following:
1. Confirm that the GPIB address you assigned to the instrument is unique and between the range
of 0 to 30. It should not be 0 or 21 because they are common controller addresses.
2. Check cabling connection. GPIB cables are heavy and can fall out of the connectors if they are
not screwed in securely.
3. Substitute cables to verify cable integrity. For example, if you can send and receive ASCII text,
but you cannot do a binary transfer, check your program and the decoding of the binary data. If
that does not resolve the problem, try another cable. ASCII text only uses seven data lines in the
cable; the binary transfer requires all eight lines.
How do I change the command set?
You can change the command set that you use with the Model 2450. The remote command sets that
are available include:
SCPI: An instrument-specific language built on the SCPI standard.
TSP: A programming language that can be used to send individual commands or combine
commands into scripts.
SCPI 2400: Allows you to run code developed for earlier Series 2400 instruments.

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
BrandKeithley
Model2450
CategoryTest Equipment
LanguageEnglish

Related product manuals