Section
6: SCPI command reference Model 2450 Interactive SourceMeter® Instrument
Reference Manual
6-106 2450-901-01 Rev. B/September 2013
Details
When you select RST, the instrument restores settings to their default values when the instrument is powered on.
When you select a SAV option, the settings in the selected saved setup are applied when the instrument is
powered on. The settings are saved using the *SAV command.
Example
SYST:POS SAV1
Set the instrument to restore the settings
that are saved in the stored setup 1 when
the instrument is powered on.
Also see
*SAV (on page 6-2)
:SYSTem:TIME
This command sets the absolute time of the instrument.
Type Affected by Where saved Default value
Command and query
Not applicable
Nonvolatile memory
See Details
Usage
:SYSTem:TIME <year>, <month>, <day>, <hour>, <minute>, <second>
:SYSTem:TIME <hour>, <minute>, <second>
:SYSTem:TIME?
:SYSTem:TIME? 1
<year>
Year; must be more than 1970
<month>
Month (1 to 12)
<day>
Day (1 to 31)
<hour>
Hour in 24-hour time format (0 to 23)
<minute>
Minute (0 to 60)
<second>
Second (0 to 60)
Details
When queried without a parameter, this command returns the present timestamp value in seconds since January
1, 1970 to the nearest second.
If you query with 1, this command returns the present timestamp in the format:
<weekday> <month> <day> <hour>:<minute>:<second> <year>
Where <weekday> is the day of the week.
Internally, the instrument bases time in UTC time. UTC time is specified as the number of seconds since Jan 1,
1970, UTC. You can use UTC time from a local time specification, or you can use UTC time from another source
(for example, your computer).
If you use os.time() but do not specify the time (hour, minute, and second) options, they default to noon for
that day. When called without a parameter (the first form), the function returns the current time.
Example
syst:time 2014, 5, 12, 5, 51, 30
Set the system time to May 12, 2014 at
05:51:30.
Also see
None

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
Model2450
CategorySourceMeter
Voltage Source Range±200 V
Current Source Range±1 A
DC Voltage200 V
DC Current1 A
Power Output20 W
Resolution6.5 digits
Display5-inch capacitive touchscreen
Operating Temperature0°C to 50°C
ConnectivityUSB, LAN, GPIB
Dimensions213 mm × 88 mm × 370 mm