Model 2450
Interactive SourceMeter® Instrument Reference Manual Section 6:
SCPI command reference
2450-901-01 Rev. B/September 2013 6-111
:TRACe:DELete
This command deletes a user-defined reading buffer.
Type Affected by Where saved Default value
Command only Not applicable Not applicable Not applicable
Usage
:TRACe:DELete <readingBuffer>
<readingBuffer>
A string that contains the name of the user-defined reading buffer to delete
Details
You cannot delete the default reading buffers, defbuffer1 and defbuffer2.
Example
TRAC:DEL "testData"
Delete the testData buffer.
Also see
Reading buffers (on page 3-10)
Remote buffer operation (on page 3-27)
:TRACe:MAKE (on page 6-114)