Section
6: SCPI command reference Model 2450 Interactive SourceMeter® Instrument
Reference Manual
6-64 2450-901-01 Rev. B/September 2013
Details
Use this command to recall the settings stored in a specific configuration point on a specific configuration list. If
you do not specify a point when you send the command, it recalls the settings stored in the first configuration
point on the specified configuration list.
Examples
:SOURce:CONF:LIST:REC "MySourceList", 5
Recalls configuration point 5 in a configuration list
named MySourceList.
:SOURce:CONF:LIST:RECall "MySourceList"
Since a point was not specified, this command
recalls configuration point 1 from a configuration
list named MySourceList.
Also see
Configuration lists (on page 3-33)
:SOURce[1]:CONFiguration:LIST:CREate (on page 6-61)
:SOURce[1]:CONFiguration:LIST:SIZE?
This command returns the size (number of configuration points) of a source configuration list.
Type Affected by Where saved Default value
Query only
Not applicable
Not applicable
Not applicable
Query
:SOURce[1]:CONFiguration:LIST:SIZE? <name>
<name>
A string that represents the name of a source configuration list
Details
This command returns the size (number of configuration points) of a source configuration list.
Examples
:SOURce:CONFigure:LIST:SIZE?
"MySourceList"
Returns the number of configuration points in a
source configuration list named
MySourceList
.
Also see
Configuration lists (on page 3-33)
:SOURce[1]:CONFiguration:LIST:CREate (on page 6-61)

Table of Contents