Interactive SourceMeter® Instrument Reference Manual Section 6:
2450-901-01 Rev. B/September 2013 6-55
[:SENSe[1]]:<function>:RELative
This command specifies the relative offset value for the measurement.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
0.0
Usage
[:SENSe[1]]:<function>:RELative <n>
[:SENSe[1]]:<function>:RELative?
The measurement function to which this setting applies:
• Current: CURRent[:DC]
• Resistance: RESistance
• Voltage:
The relative offset value:
• Current: −1 to 1
• Resistance: –200000000 to 200000000
•
−
Details
This command specifies the relative offset value that is used for measurements. When relative offset is enabled,
all subsequent measured readings are offset by the value that is set for this command.
Example
CURR:REL .5
CURR:REL:STAT ON
Set the relative offset for current
measurements to 0.5. Enable relative offset.
Also see
Relative offset (on page 3-1)
[:SENSe[1]]:<function>:RELative:ACQuire (on page 6-56)
[:SENSe[1]]:<function>:RELative:STATe (on page 6-57)