8: TSP command reference Model 2450 Interactive SourceMeter® Instrument
Reference Manual
8-90 2450-901-01 Rev. B/September 2013
smu.measure.autozero.once()
This function causes the instrument to autozero once.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
smu.measure.autozero.once()
Details
This command forces a refresh of the reference and zero measurements that are used for the present aperture
setting.
When autozero is set to off, the instrument may gradually drift out of specification. To minimize the drift, you can
send the once command to make a reference and zero measurement immediately before a test sequence.
Example
smu.measure.autozero.once()
Forces a refresh of the reference and zero
measurements.
Also see
smu.measure.autozero.enable (on page 8-89)

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
Model2450
CategorySourceMeter
Voltage Source Range±200 V
Current Source Range±1 A
DC Voltage200 V
DC Current1 A
Power Output20 W
Resolution6.5 digits
Display5-inch capacitive touchscreen
Operating Temperature0°C to 50°C
ConnectivityUSB, LAN, GPIB
Dimensions213 mm × 88 mm × 370 mm

Related product manuals