EasyManuals Logo
Home>Keithley>Test Equipment>2450

Keithley 2450 User Manual

Keithley 2450
878 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #621 background imageLoading...
Page #621 background image
Section
8: TSP command reference Model 2450 Interactive SourceMeter® Instrument
Reference Manual
8-90 2450-901-01 Rev. B/September 2013
smu.measure.autozero.once()
This function causes the instrument to autozero once.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
smu.measure.autozero.once()
Details
This command forces a refresh of the reference and zero measurements that are used for the present aperture
setting.
When autozero is set to off, the instrument may gradually drift out of specification. To minimize the drift, you can
send the once command to make a reference and zero measurement immediately before a test sequence.
Example
smu.measure.autozero.once()
Forces a refresh of the reference and zero
measurements.
Also see
smu.measure.autozero.enable (on page 8-89)

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
BrandKeithley
Model2450
CategoryTest Equipment
LanguageEnglish

Related product manuals