Interactive SourceMeter® Instrument Reference Manual Appendix C:
Status model
2450-901-01 Rev. B/September 2013 C-5
Programmable status register sets
You can program the registers in the Questionable Event Register and Operation Event Register
sets.
These event registers contain bits that identify the state of an instrument condition or event. They also
contain bits that determine if those events are sent to the Status Byte Register. You can enable the
events which cause the associated bit to be set in the Status Byte Register.
The Questionable and Operation Event Registers are identical except that they set different bits in the
Status Byte Register. The Questionable Event Registers set both the MSB and QSM bits; the
Operation Event Registers set the OSB bit.
Each 16-bit register set includes the following registers:
Condition: A read-only register that is constantly updated to reflect the present operating
conditions of the instrument. You can determine which events set or clear the bits.
Event: A read-only register that sets a bit to 1 when an applicable event occurs. The bit remains
at 1 until the register is reset. This register is reset when power is cycled or when a *CLS
command is sent. You can determine which events set the bits.
Event enable: A read-write register that determines which events set the summary bit in the
Status Byte Register. For example, if a bit is a 1 in the event register and the corresponding bit is
a 1 in the Event Enable Register, bits in the Status Byte Register are set. If the event enable bit is
set in the Questionable Event Registers, the event sets the MSB and QSM bits in the Status Byte
Register. If the event enable bit is set in the Operation Event Registers, the event sets the OSB
bit in the Status Byte Register.
On power-up, all bits in the Questionable Event and Operation Event registers are set to 0.

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
Model2450
CategorySourceMeter
Voltage Source Range±200 V
Current Source Range±1 A
DC Voltage200 V
DC Current1 A
Power Output20 W
Resolution6.5 digits
Display5-inch capacitive touchscreen
Operating Temperature0°C to 50°C
ConnectivityUSB, LAN, GPIB
Dimensions213 mm × 88 mm × 370 mm

Related product manuals