Model 2450
Interactive SourceMeter® Instrument Reference Manual Section 6:
SCPI command reference
2450-901-01 Rev. B/September 2013 6-67
:SOURce[1]:<function>:DELay:AUTO
This command enables or disables the autodelay that occurs when the source is turned on.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Source configuration list
Save settings
Source configuration list
1 (ON)
Usage
:SOURce[1]:<function>:DELay:AUTO <state>
:SOURce[1]:<function>:DELay:AUTO?
<function>
The source function to which this setting applies:
Current: CURRent
Voltage:
VOLTage
<state>
Disable the source auto delay: ON or 0
Enable the source auto delay: OFF or 1
Details
When auto delay is turned on, the actual delay that is set depends on the range.
When source autodelay is on, if you set a source delay, the autodelay is turned off.
Example
SOUR:CURR:DEL:AUTO OFF
Turn off auto delay when current is being
sourced.
Also see
:SOURce[1]:<function>:DELay (on page 6-66)

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
Model2450
CategorySourceMeter
Voltage Source Range±200 V
Current Source Range±1 A
DC Voltage200 V
DC Current1 A
Power Output20 W
Resolution6.5 digits
Display5-inch capacitive touchscreen
Operating Temperature0°C to 50°C
ConnectivityUSB, LAN, GPIB
Dimensions213 mm × 88 mm × 370 mm

Related product manuals