Example
reset()
testData = buffer.make(50)
trigger.model.load("SimpleLoop", 3, 0, testData)
trigger.model.initiate()
waitcomplete()
print(testData.relativetimestamps[1])
printbuffer(1, 3, testData.relativetimestamps)
Create a reading buffer named
testData, configure the instrument to
make three measurements, and store
the readings in the buffer.
Print the relative timestamp for the first
reading in the buffer.
Example output:
0
Print the relative timestamp for the
reading 1 through 3 in the buffer.
Example output:
0, 0.383541, 0.772005
Also see
buffer.delete() (on page 8-9)
buffer.make() (on page 8-11)
bufferVar.clear() (on page 8-17)
print() (on page 8-77)
printbuffer() (on page 8-78)
Reading buffers (on page 3-10)
Remote buffer operation (on page 3-27)
bufferVar.seconds
This attribute contains the nonfractional seconds portion of the timestamp when the reading was stored in UTC
format.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (R)
Yes
Restore configuration
Instrument reset
Power cycle
Not applicable
Not applicable
Usage
nonFracSeconds = bufferVar.seconds[N]
nonFracSeconds
The nonfractional seconds portion of the timestamp when the reading was stored
bufferVar
The name of the reading buffer, which may be a default buffer (defbuffer1 or
defbuffer2) or a user-defined buffer
N
The reading number N; can be any value between 1 and the number of readings in
the buffer; use the bufferVar.n command to determine the number of readings
in the buffer