3: Functions and features Model 2450 Interactive SourceMeter® Instrument
Reference Manual
3-58 2450-901-01 Rev. B/September 2013
Sweep programming examples
The following examples show programming examples of typical sweeps.
Linear sweep with a voltage source
The following examples perform a linear sweep that uses a voltage source. They perform the
following actions:
Reset the instrument to its defaults.
Set the source function to voltage.
Set the source range to 20 V.
Set the measure function to current.
Set the current range to 100 µA.
Set up a linear sweep that sweeps from 0 to 10 volts in 20 steps with a source delay of 1 ms, a
sweep count of 1, and a fixed source range. In TSP only, name the configuration list that is
created for this sweep VoltLinSweep.
Start the sweeps.
No buffer is defined, so the data is stored in defbuffer1. See Reading buffers (on page 3-10
) for
more information on reading buffers.
Using SCPI commands
*RST
SOUR:FUNC VOLT
SOUR:VOLT:RANG 20
SENS:FUNC "CURR"
SENS:CURR:RANG 100e-6
SOUR:SWE:VOLT:LIN 0, 10, 20, 1e-3, 1, FIXED
INIT
Using TSP commands
reset()
smu.source.func = smu.FUNC_DC_VOLTAGE
smu.source.range = 20
smu.measure.func = smu.FUNC_DC_CURRENT
smu.measure.range = 100e-6
smu.source.sweeplinear("VoltLinSweep", 0, 10, 20, 1e-3, 1, smu.RANGE_FIXED)
trigger.model.initiate()