Model 2450
Interactive SourceMeter® Instrument Reference Manual Section 3:
Functions and features
2450-901-01 Rev. B/September 2013 3-39
Creating, storing, and performing operations on configuration lists
and points
To create a configuration point, you need to:
Create a new configuration list and give it a name or use a specific configuration list that already
exists on the instrument
Configure the instrument with the settings that you want to store in a configuration point
Store the active settings into a configuration point on the specified configuration list
After you store configuration points to a configuration list, you can perform the following operations on
a specific configuration point:
Recall a configuration point and restore the stored settings to the active state
View the contents of a configuration point
Delete a configuration point or delete the entire configuration list
You can perform configuration lists operations from the front panel (see
Using the front panel for
configuration list operations (on page 3-39)) or by using remote commands (see Using remote
commands for configuration list operations (on page 3-43).
Using the front panel for configuration list operations
This section describes how to store the active settings to a specific point on a configuration list.
See What is a configuration point? (on page 3-35
) for information about how the instrument adds
configuration points to configuration lists before reading this section.
The following topics provide information to:
Create an example measure configuration list named MyMeasList.
Store two example configuration points on MyMeasList. The following table provides information
about specific settings for each configuration point.
View the contents of a specific configuration point.
Recall a configuration point from MyMeasList.
This example creates a measure configuration list; however, you can use the same process to create
a source configuration list.
Front panel configuration list menu overview
To display the configuration list menus from the main menu:
1. Press the MENU key.
2. Under Measure, select Config List. The MEASURE CONFIGURATION LISTS screen is
displayed. Use this menu for operations on measure configuration lists.
3. Under Source, select Config List. The SOURCE CONFIGURATION LISTS screen is displayed.
Use this menu for operations on source configuration lists.

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
BrandKeithley
Model2450
CategoryTest Equipment
LanguageEnglish

Related product manuals