Section
8: TSP command reference Model 2450 Interactive SourceMeter® Instrument
Reference Manual
8-248 2450-901-01 Rev. B/September 2013
tspnet.write()
This function writes a string to the remote instrument.
Type TSP-Link accessible
Affected by Where saved Default value
Function
No
Usage
tspnet.write(connectionID, inputString)
connectionID
The connection ID returned from
tspnet.connect()
inputString
The string to be written
Details
The tspnet.write() function sends inputString to the remote instrument. It does not wait for command
completion on the remote instrument.
The Model 2450 sends inputString to the remote instrument exactly as indicated. The inputString must
contain any necessary new lines, termination, or other syntax elements needed to complete properly.
Because tspnet.write() does not process output from the remote instrument, do not send commands that
generate too much output without processing the output. This command can stop executing if there is too much
unprocessed output from previous commands.
Example
tspnet.write(myID, "runscript()\r\n")
Commands the remote instrument to execute
a command or script named runscript() on
a remote device identified in the system as
myID
.
Also see
tspnet.connect() (on page 8-237)
tspnet.read() (on page 8-241)
upgrade.previous()
This function returns to a previous version of the Model 2450 firmware.
Type TSP-Link accessible
Affected by Where saved Default value
Function
No
Usage
upgrade.previous()
Details
This function allows you to revert to an earlier version of the firmware.
When you send this function, the instrument searches the flash drive that is inserted in the front-panel USB port
for an upgrade file. If the file is found, the instrument performs the upgrade. An error is returned if an upgrade
file is not found.
Also see
Upgrading the firmware (on page A-3)
upgrade.unit() (on page 8-249)

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
Model2450
CategorySourceMeter
Voltage Source Range±200 V
Current Source Range±1 A
DC Voltage200 V
DC Current1 A
Power Output20 W
Resolution6.5 digits
Display5-inch capacitive touchscreen
Operating Temperature0°C to 50°C
ConnectivityUSB, LAN, GPIB
Dimensions213 mm × 88 mm × 370 mm

Related product manuals