Section
6: SCPI command reference Model 2450 Interactive SourceMeter® Instrument
Reference Manual
6-168 2450-901-01 Rev. B/September 2013
:TRIGger:TIMer<n>:DELay
These commands set and read the timer delay.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Save settings
10e-6 (10 µs)
Usage
TRIGger:TIMer<n>:DELay <interval>
TRIGger:TIMer<n>:DELay?
<n>
Trigger timer number (1 to 4)
<interval>
Delay interval in seconds
Details
Each time the timer is triggered, it uses this delay period.
Reading this command returns the delay interval that will be used the next time the timer is triggered.
Example
TRIG:TIM2:DEL 50E-6
Set trigger timer 2 to delay for 50 µs.
Also see
None

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
Model2450
CategorySourceMeter
Voltage Source Range±200 V
Current Source Range±1 A
DC Voltage200 V
DC Current1 A
Power Output20 W
Resolution6.5 digits
Display5-inch capacitive touchscreen
Operating Temperature0°C to 50°C
ConnectivityUSB, LAN, GPIB
Dimensions213 mm × 88 mm × 370 mm