Section
6: SCPI command reference Model 2450 Interactive SourceMeter® Instrument
Reference Manual
6-116 2450-901-01 Rev. B/September 2013
:TRACe:SAVE
This command saves data from the specified reading buffer to a USB flash drive.
Type Affected by Where saved Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:TRACe:SAVE <fileName>, <bufferName>
:TRACe:SAVE <fileName>, <bufferName>, <timeFormat>
:TRACe:SAVE <fileName>, <bufferName>, <timeFormat>, <start>, <end>
<fileName>
A string that indicates the name of the file on the USB flash drive in which to save
the reading buffer
<bufferName>
A string that indicates the reading buffer; the default buffers (defbuffer1 or
defbuffer2) or the name of a user-defined buffer; if no buffer is specified, this
parameter defaults to defbuffer1
<timeFormat>
Defines how date and time information from the buffer is saved in the file on the
USB flash drive; the values are:
Dates, times, and fractional seconds are saved; the default value: FORMat
Relative time stamps are saved: RELative
Seconds and fractional seconds are saved: RAW
Time stamps are saved:
STAMp
<start>
Defines the starting point in the buffer to start saving data
<end>
Defines the ending point in the buffer to stop saving data

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
BrandKeithley
Model2450
CategoryTest Equipment
LanguageEnglish

Related product manuals