8-18 2450-901-01 Rev. B/September 2013
bufferVar.dates
This attribute contains the dates of readings stored in the reading buffer.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (R) Yes
Restore configuration
Instrument reset
Power cycle
Not applicable Not applicable
Usage
date = bufferVar.dates[N]
date
The date of readings stored in bufferVar element N
bufferVar
The name of the reading buffer, which may be a default buffer (defbuffer1 or
defbuffer2
) or a user-defined buffer
N
The reading number N; can be any value between 1 and the number of readings in
the buffer; use the bufferVar.n command to determine the number of readings
in the buffer
Details
This attribute contains the dates of readings stored in the reading buffer.
Example
reset()
testData = buffer.make(50)
trigger.model.load("SimpleLoop", 3, 1, testData)
trigger.model.initiate()
waitcomplete()
print(testData.dates[1])
printbuffer(1, testData.n, testData.dates)
Create a reading buffer named
testData, configure the instrument to
make three measurements, and store the
readings in the buffer.
Print the first reading date.
Example output:
03/01/2013
Prints the dates for readings 1 through the
last reading in the buffer.
Example output:
03/01/2013, 03/01/2013,
03/01/2013
Also see
buffer.delete() (on page 8-9)
buffer.make() (on page 8-11)
bufferVar.clear() (on page 8-17)
print() (on page 8-77)
printbuffer() (on page 8-78)
Reading buffers (on page 3-10)
Remote buffer operation (on page 3-27)