Model 2450
Interactive SourceMeter® Instrument Reference Manual Section 6:
SCPI command reference
2450-901-01 Rev. B/September 2013 6-125
:TRIGger:BLENder<n>:CLEar
This command clears the blender event detector and resets the overrun indicator of blender <n>.
Type Affected by Where saved Default value
Command only Not applicable Not applicable Not applicable
Usage
:TRIGger:BLENder<n>:CLEar
<n>
The blender number (1 or 2)
Details
This command sets the blender event detector to the undetected state and resets the overrun indicator of the
event detector.
Example
:TRIG:BLEN2:CLE
Clears the event detector for blender 2.
Also see
None
:TRIGger:BLENder<n>:MODE
This command selects whether the blender performs OR operations or AND operations.
Type Affected by Where saved Default value
Command and query
Recall settings
Instrument reset
Power cycle
Trigger blender clear
Save settings
AND
Usage
:TRIGger:BLENder<n>:MODE <operation>
:TRIGger:BLENder<n>:MODE?
<n>
The blender number (1 or 2)
<operation>
The type of operation:
OR
AND