8: TSP command reference Model 2450 Interactive SourceMeter® Instrument
8-30 2450-901-01 Rev. B/September 2013
Example
testData = buffer.make(50)
smu.source.output = smu.ON
trigger.model.load("SimpleLoop", 2, 0, testData)
trigger.model.initiate()
waitcomplete()
printbuffer(1, 2, testData.sourcestatuses)
Create a reading buffer named
testData, configure the instrument to
make two measurements, and store the
readings in the buffer.
Turn on the source output.
Print the source status for the readings
in testData.
Output:
128, 128
Indicating that the status is
Also see
buffer.make() (on page 8-11)
bufferVar.clear() (on page 8-17)
buffer.delete() (on page 8-9)
print() (on page 8-77)
printbuffer() (on page 8-78)
Reading buffers (on page 3-10)
Remote buffer operation (on page 3-27)
bufferVar.sourceunits
This attribute contains the units of measure of the source.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (R) Yes
Instrument reset
Not applicable Not applicable
Usage
readingUnits = bufferVar.sourceunits[N]
The units of measure of the source:
• Volt DC
•
The name of the reading buffer, which may be a default buffer (defbuffer1 or
defbuffer2
) or a user-defined buffer
The reading number N
; can be any value between 1 and the number of readings in the
buffer; use the bufferVar.n command to determine the number of readings in the