Interactive SourceMeter® Instrument Reference Manual Section 6: SCPI
2450-901-01 Rev. B/September 2013 6-137
Details
When the trigger model reaches a configuration recall next building block, the settings at the next index point in a
configuration list are restored.
Each time this block is encountered, the settings at the next index point in the configuration list are recalled and
take effect before the next step executes. When the last index point in the list is reached, it returns to the first
point.
Example
TRIG:BLOC:CONF:NEXT 12, "SOURCE_LIST"
Set trigger block 12 to restore the settings
from the next index point that is stored in the
configuration list
.
Also see
Configuration lists (on page 3-33)
:TRIGger:BLOCk:CONFig:PREVious
This command defines a trigger model block that recalls the settings stored at the previous index point in a
measure or source configuration list.
Type Affected by Where saved Default value
Instrument reset
Usage
:TRIGger:BLOCk:CONFig:PREVious <blockNumber>, <configurationList>
The sequence of the block in the trigger model
The measure or source configuration list from which to recall settings
Details
The configuration list previous index trigger block type recalls the previous index point in a configuration list. It
configures the source or measure settings of the instrument based on the settings at that index. The trigger
model executes the settings at that index before the next block is executed.
Each time the trigger model reaches a configuration list previous block, it goes backward one index point. When
the first point in the list is reached, it goes to the last index point in the configuration list.
Example
TRIG:BLOC:CONF:PREV 14, "SOURCE_LIST"
Set trigger block 14 to restore the settings
from the previous index point that is stored
in the configuration list SOURCE_LIST.
Also see
Configuration lists (on page 3-33)