Page #833 background imageLoading...
Page #833 background image
Appendix
C: Status model Model 2450 Interactive SourceMeter® Instrument
Reference Manual
C-16 2450-901-01 Rev. B/September 2013
Clearing registers
Registers in the status model can be cleared using commands or by instrument actions. When a
registered is cleared, the bits in the register are set to 0.
The event log and all registers are cleared when instrument power is cycled.
Using SCPI commands:
To clear the event registers of the Questionable Event and Operation Event Register set, and the
event log, send:
*CLS
When using the SCPI interface, this command does not affect the Questionable Event Enable or
Operation Event Enable registers.
Using TSP commands:
To clear the event enable registers of the Questionable Event, Operation Event, and Status Byte
Register sets, and clear the event log (which clears the Error Queue of the status model), send:.
*CLS
To clear the Questionable Event register, the Operation Event register sets, and the event log, send:
status.clear()
When using the TSP interface, this command does not affect the Questionable Event Enable or
Operation Event Enable register sets.

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
Model2450
CategorySourceMeter
Voltage Source Range±200 V
Current Source Range±1 A
DC Voltage200 V
DC Current1 A
Power Output20 W
Resolution6.5 digits
Display5-inch capacitive touchscreen
Operating Temperature0°C to 50°C
ConnectivityUSB, LAN, GPIB
Dimensions213 mm × 88 mm × 370 mm