8: TSP command reference Model 2450 Interactive SourceMeter® Instrument
8-34 2450-901-01 Rev. B/September 2013
Example
testData = buffer.make(50)
smu.source.output = smu.ON
trigger.model.load("SimpleLoop", 2, 0, testData)
trigger.model.initiate()
waitcomplete()
printbuffer(1, 2, testData.statuses)
Create a reading buffer named testData,
configure the instrument to make two
measurements, and store the readings in
the buffer.
Turn on the source output.
Print the source status for the readings in
testData.
Output:
128, 128
Indicating that the status is
.
Also see
buffer.make() (on page 8-11)
bufferVar.clear() (on page 8-17)
buffer.delete() (on page 8-9)
print() (on page 8-77)
printbuffer() (on page 8-78)
Reading buffers (on page 3-10)
Remote buffer operation (on page 3-27)
bufferVar.times
This attribute contains the time when the instrument made the readings.
Type TSP-Link accessible Affected by Where saved Default value
Instrument reset
Usage
readingTime = bufferVar.dates[N]
Hours, minutes, and seconds
The name of the reading buffer, which may be a default buffer (defbuffer1 or
) or a user-defined buffer
The reading number N; can be any value between 1 and the number of readings in
the buffer; use the bufferVar.n command to determine the number of readings
Details
This attribute contains the time when the instrument made the readings.