EasyManua.ls Logo

Keithley 2450

Keithley 2450
878 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section
8: TSP command reference Model 2450 Interactive SourceMeter® Instrument
Reference Manual
8-36 2450-901-01 Rev. B/September 2013
Example 2
for x = 1, 3 do printbuffer(x, x, testData.timestamps) end
For the buffer created in Example 1, print the timestamps for the readings.
Output:
03/01/2013 14:46:07.714614838
03/01/2013 14:46:08.100468838
03/01/2013 14:46:08.487631838
Also see
buffer.delete() (on page 8-9)
buffer.make() (on page 8-11)
bufferVar.clear() (on page 8-17)
print() (on page 8-77)
printbuffer() (on page 8-78)
Reading buffers (on page 3-10)
Remote buffer operation (on page 3-27)
bufferVar.units
This attribute contains the unit of measure that is stored with readings in the reading buffer.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (R)
Yes
Restore configuration
Instrument reset
Power cycle
Not applicable
Not applicable
Usage
readingUnits = bufferVar.units[N]
readingUnits
Volt DC: Voltage measurement
Ohm: Resistance measurement
Watt DC: Power measurement
Amp DC: Current measurement
%: Math is set to percent for the measurements
mX+b: Math is set to mx+b for the measurements
Reciprocal: Math is set to reciprocal for the measurements
bufferVar
The name of the reading buffer, which may be a default buffer (defbuffer1 or
defbuffer2
) or a user-defined buffer
N
The reading number N; can be any value between 1 and the number of readings in
the buffer; use the bufferVar.n command to determine the number of readings
in the buffer
Details
This attribute contains the unit of measure that is stored with readings in the reading buffer.

Table of Contents

Other manuals for Keithley 2450

Related product manuals