Model 2450
Interactive SourceMeter® Instrument Reference Manual Section 8:
TSP command reference
2450-901-01 Rev. B/September 2013 8-43
digio.line[N].mode
This attribute sets the digital I/O line to be a digital line or trigger model line and sets the line as an input, output,
or open-drain. You can also use this attribute to configure synchronous triggering modes.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes
Restore configuration
Instrument reset
Power cycle
Configuration script digio.MODE_DIGITAL_IN
Usage
lineState = digio.line[N].mode
digio.line[N].mode = lineState
lineState
Digital input: digio.MODE_DIGITAL_IN
Digital output: digio.MODE_DIGITAL_OUT
Digital open-drain: digio.MODE_DIGITAL_OPEN_DRAIN
Trigger model input: digio.MODE_TRIGGER_IN
Trigger model output: digio.MODE_TRIGGER_OUT
Trigger model open-drain: digio.MODE_TRIGGER_OPEN_DRAIN
Synchronous master: digio.MODE_SYNCHRONOUS_MASTER
Synchronous acceptor:
digio.MODE_SYNCHRONOUS_ACCEPTOR
N
The digital I/O line (1 to 6)
Details
Use this command to set the line as a digital line or as a trigger line, and configure it as an input, output, or open-
drain. A digital line allows direct control of the digital I/O lines; a trigger line controls the digital I/O lines through
trigger commands.
The following settings of lineState set the line for direct control as a digital line:
digio.MODE_DIGITAL_IN
digio.MODE_DIGITAL_OUT
digio.MODE_DIGITAL_OPEN_DRAIN
The following settings of lineState set the line as a trigger line:
digio.MODE_TRIGGER_IN
This line state uses the edge setting specified by trigger.digin[N].edge attribute.
digio.MODE_TRIGGER_OUT
digio.MODE_TRIGGER_OPEN_DRAIN
When configured as an open-drain, the line becomes an open-drain signal with a 100 k pull-up resistor. This is
true when the line is configured for direct-control or for use as a trigger line. When used as a trigger line, the line
uses the edge setting like digio.MODE_TRIGGER_IN.
When the line is set as a synchronous acceptor, the line detects the falling-edge input triggers and automatically
latches and drives the trigger line low; asserting an output trigger releases the latched line. When set as a
synchronous master, the line detects rising-edge triggers as input; for output, the line asserts a TTL-low pulse.
Example
digio.line[1].mode = digio.MODE_TRIGGER_OUT
Set digital I/O line 1 to be an output trigger
line.
Also see
trigger.digin[N].edge (on page 8-171)
Digital I/O port (on page 3-62)

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
Model2450
CategorySourceMeter
Voltage Source Range±200 V
Current Source Range±1 A
DC Voltage200 V
DC Current1 A
Power Output20 W
Resolution6.5 digits
Display5-inch capacitive touchscreen
Operating Temperature0°C to 50°C
ConnectivityUSB, LAN, GPIB
Dimensions213 mm × 88 mm × 370 mm

Related product manuals