Interactive SourceMeter® Instrument Reference Manual Section 3:
2450-901-01 Rev. B/September 2013 3-55
Front-panel sweep options
Option Description
Sets the source limit for measurements. The Model 2450 cannot source levels
that exceed this limit.
Determines if the sweep is stopped immediately if a limit is exceeded. You can
select one of the following options:
• ON: Abort the sweep if a limit is exceeded.
•
OFF: Complete the sweep even if a limit is exceeded.
The source range that is used for the sweep. You can select one of the following
options:
• Best Fixed: The instrument selects a single fixed source range that will
accommodate all the source levels in the sweep. This avoids overshoots
during sweeps.
• Auto: The instrument selects the most sensitive source range for each
source level in the sweep.
• Fixed: The source remains on the range that is set when the sweep is
started. If a sweep point exceeds the source range capability, the source
will output the maximum level for that range.
Setting up a sweep using SCPI commands
To set up a sweep using SCPI commands, you send one of the following commands:
• :SOURce[1]:SWEep:<function>:LINear: Sets up a linear sweep for a set number of
measurement points.
• :SOURce[1]:SWEep:<function>:LINear:STEP: Sets up a linear source sweep
configuration list and trigger model with a fixed number of steps.
• :SOURce[1]:SWEep:<function>:LIST: Sets up a sweep based on a configuration list, which
allows you to customize the sweep.
• :SOURce[1]:SWEep:<function>:LOG: Sets up a logarithmic sweep for a set number of
measurement points.
To create a sweep:
1. Set the source function using :SOURce[1]:FUNCtion[:MODE].
2. Set the source range using :SOURce[1]:<function>:RANGe.
3. Set any other source settings that apply to your sweep. You must set source settings before the
sweep function is called.
4. If you are using :SOURce[1]:SWEep:<function>:LIST, set up the source configuration list
for your sweep.
5. Set the parameters for the sweep command.
6. Set the measurement function using [:SENSe[1]]:FUNCtion.
7. Set the measurement range using [:SENSe[1]]:<function>:RANGe[:UPPer].
8. Make any other settings appropriate to your sweep.
9. Send :INITiate to start the sweep.
To save your settings, save them to a user-saved setup using the *SAV command.