6: SCPI command reference Model 2450 Interactive SourceMeter® Instrument
Reference Manual
6-32 2450-901-01 Rev. B/September 2013
When the zero output-off state is selected, when you turn off the output:
The measurement sense is changed to 2-wire sense
The source function is set to voltage
The source voltage is set to 0
Set the range to the presently selected range (turn off autorange)
Program the voltage DAC to zero, and the current DAC to full scale of the present current range.
When the zero output-off state is selected, you can use the instrument as an ammeter because it is outputting
0 V.
When the guard output-off state is selected and the output is turned off, the following actions occur:
The measurement sense is changed to 2-wire sense
The current source is selected and set to 0 A
The voltage limit is set to 10% full scale of the present voltage range
If you send this command without the <function> parameter, it will set the output-off state for all
functions.
Example
:OUTP:CURR:SMOD HIMP
Sets the output-off state for the current
function so that the instrument opens the
output relay when the output is turned off.
Also see
Output-off state (on page 2-82)
:OUTPut[1][:STATe] (on page 6-34)

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
Model2450
CategorySourceMeter
Voltage Source Range±200 V
Current Source Range±1 A
DC Voltage200 V
DC Current1 A
Power Output20 W
Resolution6.5 digits
Display5-inch capacitive touchscreen
Operating Temperature0°C to 50°C
ConnectivityUSB, LAN, GPIB
Dimensions213 mm × 88 mm × 370 mm

Related product manuals