2: General operation Model 2450 Interactive SourceMeter® Instrument
Reference Manual
2-20 2450-901-01 Rev. B/September 2013
The main menu is organized into six submenus, which are labeled in green across the top of the
display. The icons in each column open interactive screens.
QuickSet controls menu
The QuickSet controls menu (centered under Source and Measure on the main menu) allows you to
select predefined setups for Function and "One-Touch" instrument-type emulation. This menu also
contains a sliding scale adjustment for Performance (resolution versus speed).
When you select a Quick Setup, the instrument turns the output on. Carefully consider and configure
the appropriate output-off state, source, and limits before connecting the Model 2450 to a device that
can deliver energy, such as other voltage sources, batteries, capacitors, or solar cells. Configure the
settings that are recommended for the instrument before making connections to the device. Failure
to consider the output-off state, source, and limits may result in damage to the instrument or to the
device under test (DUT).
The table below shows the default settings for each of the quick setups.
QuickSet controls default settings
QuickSet selection Source settings Measure settings
Source Voltage and Measure
Current
Voltage 0 mV, 20 mV range,
current limit 105 µA
Current, 100 nA range, 2-wire,
autozero on
Source Voltage and Measure
Voltage
Voltage 0 mV, 20 mV range,
current limit 105 µA
Voltage, 20 mV range, 2-wire,
autozero on
Source Voltage and Measure
Resistance
Voltage 0 mV, 20 mV range,
current limit 105 µA
Resistance, 100 nA range, 2-wire,
autozero on
Source Voltage and Measure
Power
Voltage 0 mV, 20 mV range,
current limit 105 µA
Power, 100 nA range, 2-wire,
autozero on
Source Current and Measure
Voltage
Current 0 nA, 10 nA range,
voltage limit 21 V
Voltage, 20 mV range, 2-wire,
autozero on
Source Current and Measure
Current
Current 0 nA, 10 nA range,
voltage limit 21 V
Current, 10 nA range, 2-wire,
autozero on
Source Current and Measure
Resistance
Current 0 nA, 100 mA range,
voltage limit 21 V
Resistance, 2 range, 2 wire,
autozero on
Source Current and Measure
Power
Current 0 nA, 100 mA range,
voltage limit 21 V
Power, 20 mV range, 2 wire,
autozero on
Voltmeter
Current, 0 A, 10 µA range, voltage
limit 200 V, output on, continuous
triggering mode
Voltage, autorange on, NPLC = 1
Ammeter
Voltage, 0 V, 20 V range, 1 A
current limit, output on,
continuous triggering mode
Current, autorange on
Ohmmeter
Current, output on, 0.1 µA,
autorange, continuous triggering
mode
Resistance, auto-ohms,
autorange on, 2-wire or 4-wire
(user-specified)
Power Supply
Voltage, fixed source level (user-
specified), autorange on, source
readback on, current limit (user-
specified), source delay 0.1 s,
output on, continuous triggering
mode
Current, autorange on, 2-wire or
4-wire (user-specified)

Table of Contents

Other manuals for Keithley 2450

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2450 and is the answer not in the manual?

Keithley 2450 Specifications

General IconGeneral
BrandKeithley
Model2450
CategoryTest Equipment
LanguageEnglish

Related product manuals