2: General operation Model 2450 Interactive SourceMeter® Instrument
2-20 2450-901-01 Rev. B/September 2013
The main menu is organized into six submenus, which are labeled in green across the top of the
display. The icons in each column open interactive screens.
QuickSet controls menu
The QuickSet controls menu (centered under Source and Measure on the main menu) allows you to
select predefined setups for Function and "One-Touch" instrument-type emulation. This menu also
contains a sliding scale adjustment for Performance (resolution versus speed).
When you select a Quick Setup, the instrument turns the output on. Carefully consider and configure
the appropriate output-off state, source, and limits before connecting the Model 2450 to a device that
can deliver energy, such as other voltage sources, batteries, capacitors, or solar cells. Configure the
settings that are recommended for the instrument before making connections to the device. Failure
to consider the output-off state, source, and limits may result in damage to the instrument or to the
device under test (DUT).
The table below shows the default settings for each of the quick setups.
QuickSet controls default settings
QuickSet selection Source settings Measure settings
Source Voltage and Measure
Current
Voltage 0 mV, 20 mV range,
current limit 105 µA
Current, 100 nA range, 2-wire,
autozero on
Source Voltage and Measure
Voltage
Voltage 0 mV, 20 mV range,
current limit 105 µA
Voltage, 20 mV range, 2-wire,
autozero on
Source Voltage and Measure
Resistance
Voltage 0 mV, 20 mV range,
current limit 105 µA
Resistance, 100 nA range, 2-wire,
autozero on
Source Voltage and Measure
Power
Voltage 0 mV, 20 mV range,
current limit 105 µA
Power, 100 nA range, 2-wire,
autozero on
Source Current and Measure
Voltage
Current 0 nA, 10 nA range,
voltage limit 21 V
Voltage, 20 mV range, 2-wire,
autozero on
Source Current and Measure
Current 0 nA, 10 nA range,
Current, 10 nA range, 2-wire,
Source Current and Measure
Resistance
Current 0 nA, 100 mA range,
voltage limit 21 V
Resistance, 2 Ω range, 2 wire,
Source Current and Measure
Current 0 nA, 100 mA range,
Power, 20 mV range, 2 wire,
Current, 0 A, 10 µA range, voltage
limit 200 V, output on, continuous
Voltage, autorange on, NPLC = 1
Voltage, 0 V, 20 V range, 1 A
current limit, output on,
continuous triggering mode
Current, output on, 0.1 µA,
autorange, continuous triggering
autorange on, 2-wire or 4-wire
Voltage, fixed source level (user-
specified), autorange on, source
readback on, current limit (user-
specified), source delay 0.1 s,
output on, continuous triggering
Current, autorange on, 2-wire or
4-wire (user-specified)