Interactive SourceMeter® Instrument Reference Manual Appendix B:
2450-901-01 Rev. B/September 2013 B-5
*IDN?
This command retrieves the identification string of the instrument.
Type Affected by Where saved Default value
Query only None Not applicable Not applicable
Usage
*IDN?
Details
The identification string includes the manufacturer, model number, serial number, and firmware revision of the
instrument. The string is formatted as follows:
KEITHLEY INSTRUMENTS INC.,MODEL nnnn,xxxxxxx,yyyyy
Where:
• nnnn is the model number
• xxxxxxx is the serial number
• yyyyy is the firmware revision level
Example
Output:
KEITHLEY INSTRUMENTS INC.,MODEL 2450,01234567,1.0.0i
Also see
System information (on page 2-66)
*LANG
This command determines which command set is used by the instrument.
Type Affected by Where saved Default value
Usage
*LANG <commandSet>
*LANG?
The command set to be used:
• TSP
• SCPI
•
Details
The remote command sets that are available include:
• SCPI: An instrument-specific language built on the SCPI standard.
• TSP: A programming language that can be used to send individual commands or combine commands
into scripts.
• SCPI 2400: Allows you to run code developed for earlier Series 2400 instruments.