Interactive SourceMeter® Instrument Reference Manual Appendix B:
Common commands
2450-901-01 Rev. B/September 2013 B-7
*RST
This command resets the instrument settings to their default values.
Type Affected by Where saved Default value
Command only
Usage
*RST
Details
Returns the instrument to default settings, cancels all pending commands, and cancels the response to any
previously received *OPC and *OPC? commands.
Also see
reset() (on page 8-82)