Model 2450
Interactive SourceMeter® Instrument Reference Manual Appendix B:
Common commands
2450-901-01 Rev. B/September 2013 B-9
*STB?
This command gets the serial poll byte of the instrument without clearing the request service bit.
Type Affected by Where saved Default value
Query only Not applicable Not applicable Not applicable
Usage
*STB?
Details
The status byte query (*STB?) command is similar to a serial poll, but it is processed like any other instrument
command.
The *STB? command returns the same result as a serial poll, but the request service bit (bit 6) is not cleared if a
serial poll has occurred.
Example
*STB?
Queries the status byte.
Also see
None
*TRG
This command generates a trigger event from a remote command interface.
Type Affected by Where saved Default value
Command only
Usage
*TRG
Details
Use the *TRG command to generate a trigger event.
If you are using the SCPI command set, this command generates the COMMAND event. If you are using the
TSP command set, this command generates the trigger.EVENT_COMMAND event. You can use this constant
as the stimulus of any trigger object, which causes that trigger object to respond to the trigger events generated
by *TRG. See Using trigger events to start actions in the trigger model (on page 3-109
).
Also see
:INITiate[:IMMediate] (on page 6-124)