Remote Commands
R&S
®
FSW
1086User Manual 1173.9411.02 ─ 43
Example:
CORR:CVL:SEL 'LOSS_TAB_4'
Selects the conversion loss table.
CORR:CVL:MIX 'FS_Z60'
Manual operation: See "Mixer Name" on page 428
[SENSe:]CORRection:CVL:PORTs <PortType>
This command defines the mixer type in the conversion loss table. This setting is
checked against the current mixer setting before the table can be assigned to the
range.
Before this command can be performed, the conversion loss table must be selected
(see [SENSe:]CORRection:CVL:SELect on page 1086).
This command is only available with option B21 (External Mixer) installed.
Parameters:
<PortType> 2 | 3
*RST: 2
Example:
CORR:CVL:SEL 'LOSS_TAB_4'
Selects the conversion loss table.
CORR:CVL:PORT 3
Manual operation: See "Mixer Type" on page 429
[SENSe:]CORRection:CVL:SELect <FileName>
This command selects the conversion loss table with the specified file name. If
<file_name> is not available, a new conversion loss table is created.
This command is only available with option B21 (External Mixer) installed.
Parameters:
<FileName> String containing the path and name of the file.
Example:
CORR:CVL:SEL 'LOSS_TAB_4'
Manual operation: See "New Table" on page 426
See "Edit Table" on page 426
See "File Name" on page 427
[SENSe:]CORRection:CVL:SNUMber <SerialNo>
This command defines the serial number of the mixer for which the conversion loss
table is to be used. This setting is checked against the current mixer setting before the
table can be assigned to the range.
Before this command can be performed, the conversion loss table must be selected
(see [SENSe:]CORRection:CVL:SELect on page 1086).
This command is only available with option B21 (External Mixer) installed.
Parameters:
<SerialNo> Serial number with a maximum of 16 characters