Remote Commands
R&S
®
FSW
1157User Manual 1173.9411.02 ─ 43
14.8.3.2 General Marker Settings
The following commands control general marker functionality.
Remote commands exclusive to general marker functionality
DISPlay[:WINDow<n>]:MTABle......................................................................................1157
DISPlay[:WINDow<n>]:MINFo[:STATe]........................................................................... 1157
CALCulate<n>:MARKer<m>:X:SSIZe.............................................................................1157
DISPlay[:WINDow<n>]:MTABle <DisplayMode>
This command turns the marker table on and off.
Suffix:
<n>
.
irrelevant
Parameters:
<DisplayMode> ON | 1
Turns on the marker table.
OFF | 0
Turns off the marker table.
AUTO
Turns on the marker table if 3 or more markers are active.
*RST: AUTO
Example:
DISP:MTAB ON
Activates the marker table.
Manual operation: See "Marker Table Display" on page 522
DISPlay[:WINDow<n>]:MINFo[:STATe] <DisplayMode>
This command turns the marker information in all diagrams on and off.
Suffix:
<n>
.
irrelevant
Parameters:
<DisplayMode> ON | 1
Displays the marker information in the diagrams.
OFF | 0
Hides the marker information in the diagrams.
*RST: 1
Example:
DISP:MINF OFF
Hides the marker information.
Manual operation: See "Marker Info" on page 523
CALCulate<n>:MARKer<m>:X:SSIZe <StepSize>
This command selects the marker step size mode for all markers in all windows.
Analyzing Measurements (Basics)