Remote Commands
R&S
®
FSW
1372User Manual 1173.9411.02 ─ 43
Note that this command is maintained for compatibility reasons only. Limit lines no lon-
ger need to be assigned to a trace explicitly. The trace to be checked can be defined
directly (as a suffix) in the new command to activate the limit check (see
CALCulate<n>:LIMit<li>:TRACe<t>:CHECk on page 1223).
Suffix:
<n>
.
Window
<li> Limit line
<t> irrelevant
Parameters:
<TraceNumber> 1 to 6
*RST: 1
Example:
CALC:LIM2:TRAC 3
Assigns limit line 2 to trace 3.
DISPlay[:WINDow<n>]:STATe
<State>
This command changes the display state of the selected measurement window.
Note that this command is maintained for compatibility reasons only. Use the LAYout
commands for new remote control programs
(See Chapter 14.6.2, "Working with Windows in the Display", on page 1014).
Suffix:
<n>
.
Window
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
DISPlay[:WINDow<n>]:TYPE <WindowType>
This command selects the results displayed in a measurement window.
Note that this command is maintained for compatibility reasons only. Use the LAYout
commands for new remote control programs (see Chapter 14.6.2, "Working with Win-
dows in the Display", on page 1014).
The parameter values are the same as for LAYout:ADD[:WINDow]? on page 1015.
Suffix:
<n>
.
Window
Parameters:
<WindowType> DIAGram | RSUMmary | MTABle | PEAKlist | SGRam
Deprecated Commands