Remote Commands
R&S
®
FSW
1371User Manual 1173.9411.02 ─ 43
[SENSe:]ESPectrum<sb>:SBCenter.............................................................................. 1371
[SENSe:]ESPectrum<sb>:SBCount............................................................................... 1371
CALCulate<n>:LIMit<li>:TRACe<t>................................................................................1371
DISPlay[:WINDow<n>]:STATe....................................................................................... 1372
DISPlay[:WINDow<n>]:TYPE........................................................................................ 1372
HCOPy:ITEM:ALL........................................................................................................1373
SOURce<si>:EXTernal<gen>:FREQuency:SWEep[:STATe]..............................................1373
SYSTem:COMPatible................................................................................................... 1373
TRIGger[:SEQuence]:BBPower:HOLDoff........................................................................1373
TRIGger[:SEQuence]:RFPower:HOLDoff........................................................................1374
[SENSe:]ESPectrum<sb>:SBCenter <Frequency>
This command defines the center frequency of the selected sub block in a Multi-SEM
measurement.
Note that this command is maintained for compatibility reasons only. For newer remote
control programs use the [SENSe:]ESPectrum<sb>:SCENter command.
Suffix:
<sb>
.
Sub block in a Multi-SEM measurement
Parameters:
<Frequency> Frequency within the currently defined global span (see
[SENSe:]FREQuency:SPAN on page 1027 and [SENSe:
]FREQuency:CENTer on page 1025).
Range: 1 to 3
*RST: 1
Default unit: HZ
Example:
ESP2:SCENter 1GHZ
[SENSe:]ESPectrum<sb>:SBCount <Subblocks>
This command defines the number of sub blocks in the SEM measurement.
Note that this command is maintained for compatibility reasons only. For newer remote
control programs use the [SENSe:]ESPectrum<sb>:SCOunt command.
Suffix:
<sb>
.
irrelevant
Parameters:
<Subblocks> Number of sub blocks in the SEM measurement.
Range: 1 to 3
*RST: 1
Example:
ESP:SBCount 2
CALCulate<n>:LIMit<li>:TRACe<t> <TraceNumber>
This command links a limit line to one or more traces.
Deprecated Commands