Measurements and Results
R&S
®
FSW
252User Manual 1173.9411.02 ─ 43
Ref Level.....................................................................................................................253
RF Att Mode................................................................................................................253
RF Attenuation............................................................................................................ 253
Preamp........................................................................................................................254
Transducer Factor.......................................................................................................254
Limit Check <n>.......................................................................................................... 254
Abs Limit Start / Stop <n>........................................................................................... 254
Rel Limit Start / Stop <n>............................................................................................254
Multi-Limit Calc <n>.................................................................................................... 255
Min Sweep Points....................................................................................................... 255
Insert before Range / Insert after Range.....................................................................256
Delete Range.............................................................................................................. 256
Symmetrical Setup......................................................................................................256
Range Start / Range Stop
Sets the start frequency/stop frequency of the selected range.
To change the start/stop frequency of the first or last range, respectively, select the
appropriate span in the [SPAN] configuration dialog. You can set a span that is smaller
than the overall span of the ranges. In this case, the measurement includes only the
ranges that lie within the defined span and have a minimum span of 20 Hz. The first
and last ranges are adapted to the given span as long as the minimum span of 20 Hz
is not violated.
Define frequency values for each range relative to the center frequency. Center the ref-
erence range on the center frequency. The current "Tx Bandwidth" defines the mini-
mum span of the reference range (see "Channel Power Settings" on page 258).
Remote command:
[SENSe:]LIST:RANGe<ri>[:FREQuency]:STARt on page 924
[SENSe:]LIST:RANGe<ri>[:FREQuency]:STOP on page 925
Fast SEM
Activates "Fast SEM" mode for all ranges in the sweep list. For details, see Chap-
ter 7.6.4.3, "Fast SEM Measurements", on page 244.
Note: If you deactivate "Fast SEM" mode while "Symmetrical Setup" mode is on,
"Symmetrical Setup" mode is automatically also deactivated.
If you activate "Fast SEM" mode while "Symmetrical Setup" mode is on, not all range
settings can be configured symmetrically automatically.
Remote command:
[SENSe:]ESPectrum<sb>:HSPeed on page 922
Filter Type
Sets the filter type for this range.
For details on filter types, see Chapter 8.5.1.6, "Which Data May Pass: Filter Types",
on page 463.
Remote command:
[SENSe:]ESPectrum<sb>:RANGe<ri>:FILTer:TYPE on page 923
Spectrum Emission Mask (SEM) Measurement