Common Measurement Settings
R&S
®
FSW
394User Manual 1173.9411.02 ─ 43
For more information on configuring interfaces, see Chapter 13.1.1, "Remote Control
Interfaces and Protocols", on page 730.
Generator Type........................................................................................................... 394
Interface...................................................................................................................... 394
TTL Handshake...........................................................................................................394
GPIB Address/TCPIP Address / Computer Name...................................................... 394
Reference....................................................................................................................395
Edit Generator Setup File............................................................................................395
Frequency Min/ Frequency Max..................................................................................395
Level Min/ Level Max.................................................................................................. 395
Generator Type
Selects the generator type and thus defines the generator setup file to use.
For an overview of supported generators, see "Overview of Supported Generators"
on page 385. For information on generator setup files, see "Generator Setup Files"
on page 386.
Remote command:
SYSTem:COMMunicate:RDEVice:GENerator<gen>:TYPE on page 1101
Interface
Type of interface connection used.
For details on which signal generators support which interfaces, see the documenta-
tion of the corresponding signal generator.
GPIB
TCP/IP
Remote command:
SYSTem:COMMunicate:RDEVice:GENerator<gen>:INTerface on page 1100
TTL Handshake
If available for the specified generator type, this option activates TTL synchronization
via handshake.
Using the TTL interface allows for considerably higher measurement rates, because
the frequency stepping of the R&S FSW is directly coupled with the frequency stepping
of the generator.
For more information on TTL synchronization, see "TTL synchronization" on page 390.
For an overview of which generators support TTL synchronization see "Overview of
Supported Generators" on page 385.
Remote command:
SYSTem:COMMunicate:RDEVice:GENerator<gen>:LINK on page 1100
GPIB Address/TCPIP Address / Computer Name
For LAN connections: TCP/IP address of the signal generator
For GPIB connections: GPIB address of the signal generator.
Data Input and Output

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FSW Series and is the answer not in the manual?

R&S FSW Series Specifications

General IconGeneral
BrandR&S
ModelFSW Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals